AGFA CR 30-X User Manual page 117

Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

October 2006
technical
s p eCiF iCaTio N s
geNeRaL
Digitizer type
• Single cassette feed
• Throughput:
35 x 43 cm (14 x 17") = approx. 60 plates/hour
24 x 30 cm (9.5 x 12") = approx. 71 plates/hour
18 x 24 cm (7 x 9.5") = approx. 76 plates/hour
15 x 30 cm (6 x 12") = approx. 82 plates/hour
LCD display
• Machine status and error conditions
Greyscale resolution
• Data acquisition: 20 bits/pixel
• Output to processor: 16 bits/pixel
Dimensions and weight
• (W x D x H): 693 x 701 x 464 mm (27.2 x 27.6 x 18.2")
Depth including input tray: 769 mm (30.3")
• Weight: approx. 98 kg (216.05 lbs)
Power
• 220 - 240 V/50-60 Hz
Standby 120W, max 320W, 16A fuse
• 120V/60Hz (USA)
Standby 120W, max 320W, 15A fuse
• 100V/60Hz (Japan)
Standby 120W, max 320W, 15A fuse
Minimum requirements
• CR MD 4.0T General Cassette
• CR MD 4.0 General Plate
• NX
Environmental conditions
• Temperature: 15 - 30 °C (59 - 86°F)
• Humidity: 15 - 75% RH
• Magnetic fields: max. 3.8 µT in conformance with
EN 61000-4-8: level 2
• Rate of change of temperature: 0.5°C/minute (0.9°F)
Environmental effects
• Noise level: max. 65 dB (A)
• Heat dissipation: standby 120 W, max. 320 W
saFeTy
Approvals
• UL, cUL, CE
Transport details
• Temperature: -25 to +55°C (-4 to 131°F),
-25°C for max. 72 hours, +55°C for max. 96 hours
• Humidity: 5 - 95% RH
CR 30-X digitizer
3

Hide quick links:

Advertisement

Table of Contents
loading