If you notice conspicuous noise or smoke, disconnect the CR 30-X
immediately.
Do not pour water or any other liquid over the device.
The CR 30-X complies with the EN 60601-1 standard for Information
Technology. This means that, although it is absolutely safe, patients may
not come in direct contact with the equipment. Therefore the operator
console must be placed outside a radius of 1.5 m around the patient.
R = 1.5 m (1.83 m*)
Patient environment
* Deviation for UL 60601-1
Perform no other operations on the CR 30-X than those described in this
document.
Switch the system off before moving it. When reaching the new position,
switch the system on again.
Safety instructions for cleaning the CR 30-X
Pull the power plug out of the equipment prior to cleaning the equipment and
switch off the UPS, if installed.
26
Introduction
Patient environment
h= 2.5 m
(2.29 m*)
2386B EN 20060926