Table of Contents

Advertisement

3
Tuning

Calibrating the LC/TQ (Autotune)

Calibrating the LC/TQ (Autotune)

An Autotune can be run with the following ion sources: ESI and AJS ESI. The MMI
source is only supported with the Ultivo and 6475 LC/TQs. Autotune is
automatically run in both positive polarity and negative polarity.
1 If necessary, open the Method Editor window.
2 Click the TQ tab.
3 Click the Tune > Autotune section.
4 Click
5 Click
6 When the tune completes, review the report.
7 Click
Agilent 6475 Triple Quad LC/MS System User Guide
in the toolbar in the Autotune section. This button locks control of
the LC/TQ instrument. You cannot start a single sample run or a worklist
when Tune has control of the LC/TQ instrument.
in the toolbar in the LC/TQ Autotune section.
in the toolbar to release control of the LC/TQ instrument.
31