Agilent Technologies 6475 User Manual page 57

Triple quad lc/ms system
Table of Contents

Advertisement

4
Basic Operation
Setting Up and Running an Acquisition Method
To use an acquisition method that has a different data analysis (DA) method
NOTE
than the method entered in the worklist, show the column called Override DA
Method in the worklist by using the Show/Hide/Order Columns dialog box. In
this column, type the name of another method containing the DA parameters you
want to use for the sample. The DA part of this method is used instead of the DA
part of the current method.
You can also type the name of this method in the Add Multiple Samples dialog
box.
Agilent 6475 Triple Quad LC/MS System User Guide
d On the Sample Position tab, specify the sample vial locations (make sure
the specific sample tray type has been configured by right-clicking the
autosampler device image).
e Specify the locations, and click OK.
f To set up the worklist run, click Worklist Run Parameters
g On the Run Parameters tab, type the paths for the method.
h On the Data File Settings tab, enter or select the folders for the data files.
Select the File Naming options.
i
On the Additional Parameters tab, review the information and click OK.
j
To start the worklist, click Run Worklist
You run the worklist in locked or unlocked mode. When the mode is locked,
no one can change the method or the worklist while the worklist is running.
.
.
57