Agilent Technologies 6475 User Manual page 77

Triple quad lc/ms system
Table of Contents

Advertisement

5
Maintenance
Nebulizer
Installing the nebulizer
1 Insert the nebulizer part way into the spray chamber.
Do not hit the tip of the needle as you insert the nebulizer. The tip of the needle
C AU T I O N
is easily damaged.
2 Connect the nebulizer sample line to the nebulizer.
3 Turn the nebulizer clockwise and lock it in place.
4 Connect the sample nebulizer gas tubing to the nebulizer.
Do not over-tighten the LC fitting. An over-tightened fitting can crush the
C AU T I O N
tubing, which creates a restriction.
Fixing Spray Stability EMF
If spray stability EMF is active, you should inspect, adjust or flush your nebulizer
to ensure a stable spray from the CDS to the nebulizer. An unstable spray may
lead to issues with tuning and sample analysis.
1 Inspect
2 Follow the procedure to
Agilent 6475 Triple Quad LC/MS System User Guide
(page
73), adjust
necessary, replace the nebulizer needle or nebulizer
"Calibrating the LC/TQ (Autotune)"
reevaluated as part of a tune. If it within the acceptable range, the EMF
warning should be removed following completion of the tune.
(page
75), or flush
page
"Calibrating the LC/TQ (Checktune)"
on page
31. The spray stability is
65) your nebulizer. If
(page
75).
on page 30
or
77