HMI menu structure
D.1 Menu items (FW Rev. 4.00.00)
Menu
Device function,
code
Parameters
6.7.1.1
Val. uncertain
6.7.1.2
Value bad
D.1.7

Menu 7 - Sensor parameters

Menu
Device function,
code
Parameters
7.1
Sensor calib
7.1.1
Calib choice
7.1.2
App. param.
7.1.2.1
Viscosity
7.1.2.2
Correct. fact.
7.1.2.3
Fl. offset comp.
7.1.2.4
Cable len. TOFKOR Total cable length for a path.
7.1.3
Pipe geometry
7.1.3.1
Pipe diameter
7.1.3.1.
Engr.unit
1
7.1.3.1.
Pipe diameter
2
7.1.3.2
Roughness
116
Description
Error filter limit value uncertain (%)
for the average quality of the ultra‐
sonic measurement (signal pairs of
up and down). Below this value the
error status is set to uncertain and the
warning is indicated (on display and
output). The actual (maybe unstable)
measured values are displayed
Error filter value bad (%) with same
function as descripted above. Addi‐
tionally, the measured values are set
to "0" in the display and the output
signal to the error signal level
Description
Calibration choice
Application parameters
Viscosity of media
(viscosity for water at 20 °C is 0.01
cm
/s)
2
Calibration factor compensation val‐
ue for flow sensor to correct the
measured values; typical 1.0
Offset compensation value for flow
sensor (in flow unit) (in the menu
3.1.1 closed unit)
All paths must have the same length
Unit selection (m or in)
Inside diameter of pipe (m or in)
Factory setting
Setting options
Depends on system
1 %, 5 %, 10 %, 25 %,
configuration order‐
50 %, 75 %, 90 %, 95
ing, for example 50%
%, 99 %
for 2-path sensor sys‐
tem
50 %
1 %, 5 %, 10 %, 25 %,
50 %, 75, %, 90 %, 95
%, 99 %, never
Factory setting
Setting options
AUTO (SONOKIT sys‐
WET
tems)
AUTO
WET (SONO 3100
and SONO 3300 sys‐
tems)
0.01 cm
/s
0.005 to 5 cm
2
(optional unit: in
1.0
0.000001 to 100
+0.00000 m
3
/h
-max. flow to
+max.flow
(see menu 3.1.2)
Depending on selec‐
0 to 500 m
tion (factory setting
with the ordered
transducer cable
length)
m
m
in
Depending on selec‐
0.01 to 4.0 m
tion at ordering
SITRANS FUS060 with PROFIBUS PA
Operating Instructions, 09/2021, A5E02225054-AD
Read
or write
Write
Write
Read
or write
Write
/s
Write
2
/s)
2
Write
Write
Write
Write
Write