D.1.6

Menu 6 - Service

Menu
Device function,
code
Parameters
6.1
Enter code
6.2
Customer Code
6.3
Service code
6.4
Reset
6.5
Control values
6.5.1
Gain
6.5.1.1
Gain up 1
6.5.1.2
Gain dw 1
6.5.1.3
Gain dw 2
6.5.1.4
Gain up 2
6.5.1.5
Gain dw 3
6.5.1.6
Gain up 3
6.5.1.7
Gain dw 4
6.5.1.8
Gain up 4
6.5.1.9
Gain limit
6.5.2
Amplitudes
6.5.2.1
Ampl dw 1
6.5.2.2
Ampl dw 1
6.5.2.1
Ampl dw 2
6.5.2.2
Ampl dw 2
6.5.2.3
Ampl dw 3
6.5.2.4
Ampl dw 3
6.5.2.5
Ampl dw 4
6.5.2.6
Ampl dw 4
SITRANS FUS060 with PROFIBUS PA
Operating Instructions, 09/2021, A5E02225054-AD
Description
Entering the 4 digit code selected in
menu 6.2 in order to change user pa‐
rameters
Selection of a personal code.
Code 0: User parameter are not pro‐
tected. Code>0: User parameters are
protected under menu 6.
Only for service in the Siemens facto‐
ry
Re-start of unit without change of pa‐
rameters
Gain of first path up; optimal value:
40 to 100
Gain of first path down; optimal val‐
ue: 40 to 100
Gain of second path up; optimal val‐
ue: 40 to 100
Gain of second path down; optimal
value: 40 to 100
Gain of third path up
Gain of third path down
Gain of fourth path up
Gain of fourth path down
A warning level for application con‐
trol and diagnostics purposes
Amplitude of first path up
Amplitude of first path down
Amplitude of first path up
Amplitude of first path down
Amplitude of first path up
Amplitude of first path down
Amplitude of first path up
Amplitude of first path down
D.1 Menu items (FW Rev. 4.00.00)
Factory setting
Setting options
0
0 to 9999 According
to setting in menu 6.2
0
0 to 9999
Cancel
Cancel
Reset
0 to 255
0 to 255
0 to 255
0 to 255
0 to 255
0 to 255
0 to 255
0 to 255
Depends on ordered
1 to 255
sensor size
0 to 255;
optimum: 95 to 105
0 to 255;
optimum: 95 to 105
0 to 255;
optimum: 95 to 105
0 to 255;
optimum: 95 to 105
0 to 255;
optimum: 95 to 105
0 to 255;
optimum: 95 to 105
0 to 255;
optimum: 95 to 105
0 to 255;
optimum: 95 to 105
HMI menu structure
Read
or write
Write
Write
Write
Write
Read
Read
Read
Read
Read
Read
Read
Read
Write
Read
Read
Read
Read
Read
Read
Read
Read
113