6.6

Main menu: Service (menu 6)

Enter code, menu 6.1
Programming of the device by unauthorized persons can be disabled with a customer specific
code (range 0...9999) in the operating and display panel. By entering the appropriate code in
this menu item, the programming is enabled and the device settings can be changed.
-
Note
-
-
-
Customer code, menu 6.2
The personal code with which programming of the device can be enabled is entered or changed
in this menu item. The code is factory set to 0. The menu item is only accessible after entering
the personal code even when programming has already been enabled.
Service code, menu 6.3
The calibration data of the device are protected by a special code (factory code). The
appropriate menu items are available only after entering the service code. In normal application
cases, keying in the factory code is not necessary.
Reset, menu 6.4
A reset of the unit can be made without change of parameters. (Warm restart).
Totalizers will not be erased.
Control values, menu 6.5
In this menu, device-internal data used mainly for error diagnosis for max 4-tracks can be read.
The values displayed in the individual menu depend on the respective application (medium).
The following data are available:
-
-
-
-
-
-
-
-
42
Customer code (from factory the customer code is set to "0").
If the
or
key is actuated in the parameter setting level with the programming disabled,
the request for "Enter code" appears in the device display (when user code #0).
The programming is disabled after returning to the display mode after 10 min (when user
code #0).
Programming can be purposely disabled by entering any number, not the code number, in
this menu item (when user code #0).
Attention!
The programming is always enabled when the code = 0.
Gain
Amplitudes
Trigger level
Error %
Time of flight - up
Time of flight - down
Amplitude
Difference in time of flight (TOF)
SFIDK.PS.029.E5.02 / SW 2.01.07

Advertisement