Calibrating The Compression Unit - Siemens MAMMOMAT Novation DR Parts Replacement

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

32
Fig. 11: Distance

Calibrating the compression unit

Connect the service PC according to Documentation Software System Stand,
SPB7-250.816.03...
Start the service program.
In the main menu, under Configuration -> Compression, select Optimized compres-
sion. If you find the number/value 5137 in any of the (two) OPCOMP data fields shown in
the appearing mask, change it to some other arbitrary number. If the value in those data
fields differs from 5137, press F2 (save).
In the main menu, under Configuration ->Compression, select Optimized compres-
sion dr and repeat the procedure described above.
Prior to calibration, the following values must be entered in the system as default values.
Force = 0 kp
Force = 20 kp
Preset force 3 kp
Preset force 20 kp
Thickness bottom
Thickness bottom
Thickness top
Thickness top
MAMMOMAT Novation DR
0066
02D8
06
FF
0093
0
037E
180
SPB7-250.841.01.04.02
05.05
Compression Unit
Page 32 of 64
CS SD 24
0
Siemens AG
Medical Solutions