Mechanical works and adjusting

7Grid

General

These instructions describe how to disassemble the grid and mount a new one. It then
describes how to perform subsequent required tasks, mainly checks of several adjust-
ments.

Documents

The type of documentation depends on the acquisition workstation in use:
DROC
- Installation Instructions and Start-up (DROC)
- QC Manual (DROC)
WH AWS
- Start-up, WH AWS VAXXX(XXX - Depends on the installed software version)
- Software, WH AWS VAXXX (xxx - Depends on the installed software version)
- QC Manual (WH AS)
NOTE

Disassembling the grid

The disassembly of the grid is described in the instructions " Replacement of parts"
SPB7-250.841..., Repl.. of the
MAMMOMAT Novation DR
Note: removing the grid requires certain mech. adjustments. After
assembling the grid, image quality should be checked according
to
the
instructions
Start-up...,SPB7-250.815...for systems with WH AWS or Mammo-
mat Novation, Installation Instructions... SPB7-250.812...for sys-
tems with DROC.
FFDM(Disassembling the old detector / p.