Siemens MAMMOMAT Novation DR Parts Replacement page 31

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Compression Unit
Fig. 10: Church
Insert the compression plate into the compression unit. Manually move the compres-
sion plate downward toward the detector. It is important for the front edge of the com-
pression plate to be parallel to the carbon-fiber plate edge.
Manually move the compression plate upward. There must be 285 mm between the
carbon-fiber plate edge and the metal sheet under the guide for the compression plate
holder
Securely tighten the three bolts for the compression unit holder
NOTE
Siemens AG
Medical Solutions
(Fig. 11 / p.
32).
This document describes an initial calibration. Therefore, it de-
scribes cases in which a Mammomat system is equipped with a
newly mounted compression unit, and therefore either the relevant
data from another (e.g. the formerly mounted) compression unit
stored in the system or no meaningful compression calibration
data stored in the system.
SPB7-250.841.01.04.02
05.05
(1/Fig. 9 / p.
MAMMOMAT Novation DR
Page 31 of 64
CS SD 24
31