Siemens MAMMOMAT Novation DR Parts Replacement page 43

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Collimator and D814
Fig. 18: Tube assembly with collimator
Turn the system OFF.
Remove the covers from the swivel arm.
Fig. 19: Tube assembly support
Cut the cable tie
Remove the cables between the collimator and D805. For this, unplug connectors
X894/X895 on board D805 and connector X832/X837 on board D814
Remove the 4 screws M5x14
can be removed from the tube assembly support
Siemens AG
Medical Solutions
(3/Fig. 19 / p.
((1/Fig. 18 / p. 43)
SPB7-250.841.01.04.02
05.05
43).
and
(Fig. 20 / p.
Page 43 of 64
CS SD 24
(Fig. 19 / p.
(1/Fig. 19 / p.
43)). The collimator
44).
MAMMOMAT Novation DR
43
43).