Calibrating the area dose product measurement device

NOTE

CONSTANT parameter (Qt * Kg) = T

Read and record the
measurement device.

Programming

Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
Start the service program.
Select the Adjustment, Parameter menu.
In the Combo box "Parameter Groups:", select "Diamentor" .
Click on "Get from Unit".
The Parameter "1. Diamentor Existing" must be on YES .
Select parameter "2. Constant (Qt * Kg)" .
In "Value Actual:"field, enter the parameter
Click on "Set Value".
Click on "Put to Unit" .
Click on "Save to File".
NOTE
Siemens AG
Medical Solutions
This adjustment is used to calibrate the area dose product mea-
surement device of the SIREMOBIL Iso-C display.
T
value shown on the electronics unit of the area dose product
w
You must exit the service program with "Logoff" and "Quit" to
store the modified parameters to disk as a new file.