Starting Up The Mo Unit; Installation Of Software On Opdima Workstation - Siemens MAMMOMAT Novation DR Installation And Startup

Opdima sp
Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

6 - 2
Performing of DHHS maintenance, see Page 6 - 3
NOTICE
Starting up the MAMMOMAT
Switch on the MAMMOMAT.
Starting up the workstation and login
The workstation is started according to instructions in Starting up the workstation and
login in Supplement to the Instructions for Use MAMMOMAT Novation
Use the user name "service" and the password obtained from Siemens service center
when entering the system.
NOTE
To obtain optimal picture, adjust the monitor settings as follow:
Brightness: 50
Contrast: 50
Backlight: 70

Starting up the MO unit

1.
2.

Installation of software on OPDIMA WORKSTATION

The software is installed in the factory before delivery. If the software for any reason needs
to be reinstalled, refer to chapter "Reinstallation of software" in MAMMOMAT 1000/3000
Nova - Opdimar Service Instructions.
Opdima
Installation
The Opdimar external diaphragm must be used for tests and cali-
brations involving radiation.
Any adjustments of the monitor settings shall be performed by
qualified personell who are employed by Siemens or one of its
affiliates or who are otherwise autorized by Siemens or one of its
affiliates to provide such services.
Fig. 1 Magneto-optical drive (SONY)
Insert the MO disk, type 9.1 GB MO disk
For switch settings and SCSI ID, see Paragraph " Installing the MO unit" in
Chapter "Removal and replacement of sub-assemblies" in
MAMMOMAT 1000/3000 Nova - Opdimar Service Instructions.
SPB7-230.814.01
Rev. 01
10.04
Final procedures
DR
- Opdimar.
Page 2 of 4
CS PS 24
6
6
6
6
Siemens AG
Medical Solutions