Replacing basic unit components
•
Collimator check
- DROC workstation
Perform checks as instructed in the Collimator Adjustments chapter of the
appropriate Installation Instructions.
- WH AWS (syngo)
Perform checks as instructed in the Collimator Adjustments chapter of the
appropriate Startup instructions.
•
Grid line check
- DROC workstation
Checks must be performed according to the Checking Grid Lines chapter of the
appropriate Installation Instructions.
- WH AWS (syngo)
Checks must be performed according to the Image Quality / Checking grid lines
chapter of the appropriate Startup instructions.
•
System calibration
- DROC workstation
Checks must be performed according to the Image Quality chapter of the
appropriate Installation Instructions.
- WH AWS (syngo)
Checks must be performed according to the Image Quality / Checking Image
Quality chapter of the appropriate Startup instructions.
•
AEC calibration
The full AEC calibration must be performed, including the Dark Offset calibration.
- DROC workstation
Checks must be performed according to the Appendix / AEC Calibration chapter of
the corresponding Installation Instructions.
- WH AWS (syngo)
Checks must be performed according to the AEC Calibration Tool chapter of the
appropriate Software manual.
•
"H" and "D" dose adjustments
- DROC workstation
Checks must be performed according to the Checking Dose Settings / Checking
the AEC Dose settings chapter of the appropriate Installation Instructions.
- WH AWS (syngo)
Checks must be performed according to the Checking Dose Settings / Checking
the AEC Dose settings chapter of the corresponding Startup instructions.