Metal Curtain; Measurements With The Oscilloscope; Service Software; Delay Times Between Two Exposures - Siemens MAMMOMAT Novation DR Parts Replacement

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Prerequisites

Metal curtain

WARNING

Measurements with the oscilloscope

WARNING

Service software

The service PC is required for the MAMMOMAT stand service software.
NOTE

Delay times between two exposures

Delay times listed below must be observed to prevent the tube from overheating.
Exposure mAs value
max 100
max 200
max 300
max 400
max 500
Siemens AG
Medical Solutions
The edges of the metal curtain on the stand are very sharp.
They may cause severe injury.
Apply the protective strips after removing the covers
from the stand. Remove the protective strips only when
the covers are to be mounted or when vertical adjust-
ment of the swivel-arm system is necessary.
Under no circumstances should the existing ground conductor in
the mains cable be disconnected when operating the oscilloscope.
Life-threatening electric shock hazard exists.
Use the differential amplifier to take measurements for
which any resulting ground loop may falsify the result
(difference measurement).
When the generator is switched off and the service PC connected,
wait approximately 5 seconds before switching it on again.
Delay time between two exposures (seconds)
min 15
min 30
min 45
min 60
min 75
SPB7-250.841.01.04.02
05.05
MAMMOMAT Novation DR
Page 7 of 64
CS SD 24
7
0
0
0
0