Pull out the lamp holder carefully. Be careful with the lamp power supply cable.
Fig. 26: Lamp holder
•
Replace the lamp
NOTE
•
Reinstall the lamp holder
NOTE
•
Reinstall the plate with the screws
•
Make an exposure to check the X-ray/light field
•
If necessary, adjust the light field according to System Start-up with WH AWS,
SPB7-250.815... the Collimator Adjustment and Collimator Field Light Calibration
chapters.
Final steps
•
For information on backing up the stand data, see Software System Stand,
SPB7-250.816.03..., Adjustment and Service Programs.
•
Install the covers of the swivel arm.
MAMMOMAT Novation DR
(1/Fig. 26 / p. 48)
Do not touch the lamp with your fingers. Wear gloves or use a soft
cloth when replacing the lamp.
Be careful not to damage the filter disk.
SPB7-250.841.01.04.02
05.05
Collimator and D814
(1/Fig. 24 / p. 47)