Siemens MAMMOMAT Novation DR Parts Replacement page 17

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Replacing basic unit components
11. Rename MPabcd_final.map to MPabcd.map.
In File Manager, simply click the file name and change it from MPabcd_final.map to
MPabcd.map.
12. The MP in the file names must be in upper case letters. The file names on older CDs
were in lower case letters only. You must rename such files to e.g. MPabcd.cfg, MPab-
cd.map and MPabcd.smj. In File Manager, click the file name and replace the lower
case mp with the upper case MP.
Repeat the procedure for each file, so that all contain an upper case MP. Select
(highlight) and rename all three files.
13. For each file, select File -> Properties and assign it write permission.
14. Close File Manager with exit.
15. Close all desktop windows.
16. Reboot the entire system.
17. In the log-in window, log in as apps and enter the appropriate password.
18. Verify that the code dr appears on the MAMMOMAT Novation
cates that "digital radiography" mode is ready.
For WH AWS (syngo) on site:
1. Switch the MAMMOMAT Novation
First turn on the main power switch to initiate the BRICK startup.
After 5 minutes, turn on the control console.
After an additional 3 minutes, turn on the workstation.
2. On the desktop, you need to open/start Windows Explorer. To do so, first open the
syngo service software by selecting Options > Service > Local Service in the window
menu header.
3. Then select Utilities > Escape to OS.
4. Enter the following command:
start explorer
A Windows XP Explorer window opens.
5. On the C: partition, select the directory:
C:\aws\drul\data
It is recommended that the detector-specific files of the detectors used on each
MAMMOMAT Novation
uploading the detector-specific files of the new detector from the CD-ROM to the
workstation, the files for the removed detector must be stored.
6. In the directory C:\aws\drul\data, create a new (sub)directory named: MPwxyz ,
where wxyz are the four digits of the serial number of the old detector.
Siemens AG
Medical Solutions
DR
system be kept on the hard disk of the workstation. Before
SPB7-250.841.01.04.02
05.05
DR
system ON.
Page 17 of 64
CS SD 24
DR
console, which indi-
MAMMOMAT Novation DR
17