26
3. Select Modify settings.
Fig. 8:
4. Enter the values from
An example for entering the value under Modify low dose scaling is shown.
Scroll down to Modify low dose scaling.
- select save in the Duration of Change drop-down menu and
- enter the values in the Low Dose Scaling Value field and click Change Low Dose
Scaling to save it.
Make sure the values are separated by commas, e.g. 80,70,45.
5. Set the Brick AEC Array Dark Offset and the Brick AEC Array Mag Offset to the val-
ue 0.
Setting the time zone
1. Select Brick -> Time zone.
The current time zone is listed on the left of the screen.
MAMMOMAT Novation DR
Replacing basic unit electrical assemblies
BRICK - Modify settings
(Tab. 1 / p.
SPB7-250.841.01.04.02
05.05
24).
Page 26 of 64
CS SD 24
Siemens AG
Medical Solutions