Siemens MAMMOMAT Novation DR Parts Replacement page 24

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

24
2. To view the BRICK AEC configuration information, click the Brick AEC Configuration
link on the BRICK Control Panel page or the AEC link on any of the other PXCM pages.
The Brick AEC - Status page is displayed, as shown in
Fig. 7:
3. Note the corresponding effective values in
Tab. 1
BRICK values
Brick AEC High Dose Scaling Factor
Brick AEC Low Dose Scaling Factor
Dismount the BRICK
Turn OFF the entire system by switching OFF the mains input switch S700.
Remove the service cover at the back of the MAMMOMAT.
Remove all connectors and cables from the BRICK.
Remove the BRICK. Be aware of the weight of the BRICK and exercise appropriate
caution.
MAMMOMAT Novation DR
Replacing basic unit electrical assemblies
BRICK status
Item
SPB7-250.841.01.04.02
05.05
(Fig. 7 / p.
(Tab. 1 / p.
24).
Value
Page 24 of 64
CS SD 24
24).
Siemens AG