Disassembling Board D814 - Siemens MAMMOMAT Novation DR Parts Replacement

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

44

Disassembling board D814

Fig. 20: Collimator
Fig. 21: D814
Pos. 1
Pos. 2
Pos. 3
Pos. 4
Pos. 5
Pos. 6
Remove the shielding plate from board D184
Loosen connectors X833, X834, X 835, X836 and X838
Loosen board D814 from the 4 plastic holders
Board D814 can be removed from the collimator.
MAMMOMAT Novation DR
X838
X837
X836
X835
X834
X833
SPB7-250.841.01.04.02
05.05
Collimator and D814
(1/Fig. 20 / p.
44).
(Fig. 21 / p.
(1/Fig. 22 / p.
45).
Page 44 of 64
CS SD 24
0
44).
Siemens AG
Medical Solutions