Electrical Connection Of The Machine - AGFA Classic E.O.S. Installation Planning

Hide thumbs Also See for Classic E.O.S.:
Table of Contents

Advertisement

DD+DIS303.03E
8.8
Edition 6, Revision 0

Electrical connection of the machine

Power connection
Fuse protection
Leakage current
towards PE
Protective earth
Installation regulations
Mains connection in
the installation room
Required connection
cable
(scope of delivery)
Main breaker
Classic E.O.S. / Classic E.O.S. CL
(Type 5270 / 100 / 105)
1N~ 230 – 240 V; 50/60 Hz (200 - 240 V)
15 / 16 A
A GFI switch: (I
= 30 mA in compliance with
N
VDE 664) is integrated in the machine.
< 3.5 mA
< 0.1 Ω towards ground
VDE Electrical installations in the installation
room must be in compliance with the
regulations IEC 364, VDE 0100 and VDE
0107.
UL
Electrical installations in the installation
room must be in compliance with the
regulation "National Electrical Code" (NEC)
(NFPA 70).
VDE Double shockproof outlet according to DIN
49441 and CEE 7 standard plate VII
UL
Outlet for three-prong plugs
NEMA 6-20 R
VDE PVC line H05VV - F 3G 1.5 (3 wires)
cable length
usable length
UL
Cable included in the accessory box to
replace the VDE cable.
PVC line SJT 3 x AWG 12
(with plug NEMA 6-20 P)
cable length
usable length
Upon machine installation it must be ensured that
either the mains connector or an all-pole circuit
breaker for the installation on site is located close
to the machine and easily accessible.
Installation Planning
3.5 m
2.3 m
3.4 m
2.4 m
Chapter 14 / 31