Exhaust Connection Through The Floor - AGFA Classic E.O.S. Installation Planning

Hide thumbs Also See for Classic E.O.S.:
Table of Contents

Advertisement

Installation Planning
8.7.2
Chapter 14 / 30

Exhaust connection through the floor

Dimensions in mm
Figure 28
The film processor has an integrated exhaust unit.
Connection stub at the machine: ∅ 100 mm
Max. length of the exhaust hose: 5 m
If the hose is longer install an additional fan!
Exhaust volume:
min. 50 m³/h – max. 100 m³/h
For functional reasons the exhaust connection must always lead out of the
machine.
Classic E.O.S. / Classic E.O.S. CL
(Type 5270 / 100 / 105)
DD+DIS303.03E
Flexible exhaust hose
∅ 100 mm
Exhaust connection through the
floor
Exhaust connection:
2 screw-type exhaust
connections;
center hole through the floor
>
_ 114 mm
Order no.: CM+9522030091
Edition 6, Revision 0