Water Connection - AGFA Classic E.O.S. Installation Planning

Hide thumbs Also See for Classic E.O.S.:
Table of Contents

Advertisement

Installation Planning
8.6
Dimensions in mm (inch)
Chapter 14 / 24

Water connection

Also see Chapter 3
Water consumption Permanent replenishment:
Water pressure
Water temperature
pH value
Water conductivity
value
Regulations
The free water supply of the machine is in compliance with
the regulations of DIN 1988 / EN 1717:2001 (technical
regulations for the installation of drinking water).
1
2
527314sm.cdr
Figure 24
• Position the safety pressure hose, mounted at the dirt filter, all the way to
the shut-off valve or pressure reducer.
Safety
pressure hose:
• For safety we recommend to provide a floor drain close to the machine.
Classic E.O.S. / Classic E.O.S. CL
(Type 5270 / 100 / 105)
max. 3 liters / min (101.45 fl.oz. / min).
The water supply rate per square meter of processed
film can be changed via code.
2 bar (200 kPa) to 6 bar (600 kPa)
min. 5 °C
6.5 to 8
min. 3µS/cm
Reliable level detection in the water tank cannot be
guaranteed if this value is too low.
Shut-off valve
Clutch
3
4
Threaded fitting
Pressure reducer
(manometer),
2000
for water pressure > 6 bar
(78.7)
5
Flexible safety
6
pressure hose
7
Solenoid valve
Dirt filter
CM+9036260160
tested in compliance with DIN 57700 Part 600
2 m (78.74 inch) long
³/
" union nut
4
DD+DIS303.03E
-
not
included
in
shipment
Edition 6, Revision 0