Exhaust connection

Also see Chapter 3, Section 6

Exhaust connection through lower front panel (only Type 5270/100)

Dimensions in mm
Figure 27
The film processor has an integrated exhaust unit.
Connection stub at the machine: ∅ 100 mm
Max. length of the exhaust hose: 5 m.
If the hose is longer install an additional fan!
Exhaust volume:
min. 50 m³/h – max. 100 m³/h
For functional reasons the exhaust connection must always lead out of the
machine.
Classic E.O.S. / Classic E.O.S. CL
(Type 5270 / 100 / 105)
Installation Planning
Flexible exhaust hose ∅ 100 mm
Crossover
Connection piece for exhaust
hose