3
Turn the cassette over, holding the image plate in position with your other
hand.
4
Take away the cassette. The image plate remains lying on your hand.
Image
plate
5
When necessary, clean extreme contamination with ADC Digital Screen
Cleaner.
6
Moisten a cellulose cloth (non-fluffy) with the cleaning agent.
7
Rub the cleaner softly and evenly over the whole surface of the screen.
8
Leave the cassette with the clean screens open for approximately 10 minutes
to enable the solvent to evaporate.
9
Reassemble the cassette.
Make sure that the white side of the image plate, containing the phosphor, is
oriented towards the (black) tube side of the cassette.
Ensure that the image plate is within the flange on the inside of the
cassette. If you put the image plate into the cassette differently, e.g. if the
image plate lies partly in between the hinge of the cassette, it can be
irreparably damaged.