4
If an image plate is jammed in the post-scan unit, try to remove the jammed
image plate by pulling it carefully towards you.
If the access is too narrow, lift the jammed plate over the erase unit and remove it
through the upper section of the Digitizer.
If it is still not possible to remove the image plate, do not dismantle the
unit any further. Contact your local service organization.
5
Close the front doors.
The ADC Compact Plus will restart automatically.
After start-up, the operator main screen is displayed.
50
2231C EN 20040224