48 | WS-Manual-001, WS-Manual-T-001, WS-Manual-002, WS-Manual-T-002 | Operation
Figure 10: Wall stand position tracking disabled and enabled
The button is highlighted.
3. Adjust the wall stand height.
The X-ray tube stand is moving up or down accordingly.
Related Links
Collision indicator
Emergency stop button
Collision indicator
Systems with motorized movement have a collision indicator. The collision
indicator avoids collision of the X-ray tube head with the table.
The collision indicator will give a signal in following situations:
• The X-ray tube head is moved manually close than 30 cm to the table top
when performing an examination using the table.
• The X-ray tube head is moved manually closer than 10 cm to the table top
when performing an examination using the wall stand and the X-ray tube
head is rotated toward the wall stand.
The brake is activated and a single beep indicates the collision warning.
To further adjust the position, release the brake button and press it again.
3252C EN 20191206 1610
Note: The movement of the X-ray tube is automatically stopped
if the distance between the X-ray tube head and the table top
would become too small (less than 10 cm).
on page 48
on page 14