WS-Manual-001, WS-Manual-T-001, WS-Manual-002, WS-Manual-T-002 | Operation | 47
Tilting
To tilt the bucky, press and hold the button on the tilting handle and move the
bucky. The scale for the angle is visible at the mounting point of the bucky.
To lock the bucky into position, release the button on the tilting handle.
Topics:
•
X-ray tube stand tracks wall stand height
•
Collision indicator
X-ray tube stand tracks wall stand height
To keep constant position of the tube head unit relative to wall stand bucky
while adjusting wall stand height:
1. Set the required position of the X-ray tube stand.
The distance between the X-ray tube head and the table top must not be
less than 15 cm.
Position the X-ray tube head and the table top such that they do not collide
when the X-ray tube stand moves up or down.
2. On the tube head display, press the position tracking button.
CAUTION:
The maximum load for the bucky movement in vertical direction
is 20 kg. The bucky may slip downward when applying excessive
load.
Note: Do not move the bucky with excessive force to the end
stop positions.
Note: The bucky can be tilted to horizontal position. Do not use
the bucky as a seat.
WARNING:
Do not use position tracking while the patient is lying on the
table.
3252C EN 20191206 1610