38 | WS-Manual-001, WS-Manual-T-001, WS-Manual-002, WS-Manual-T-002 | Introduction to DR
400
All steel cables of X-ray tube stand and radio-
graphic wall stand shall be checked
All steel cables of X-ray tube stand and radio-
graphic wall stand shall be exchanged to maintain
fault-free operation and ensure safety for patient
and operator
Replacing the coin cell battery of the X-ray genera-
tor
Maintenance by the user
Check constant smooth movements
Check ease of movements
Check secure release and locking of brakes
Check functioning of operating controls
Check markers and warning signs
Warm-up of X-ray tube
Check all electric cables and connections for dam-
age or broken cables.
3252C EN 20191206 1610
CAUTION:
In case of functional defects or other deviations from normal
operation behavior the unit has to be switched off immediately
and the service to be informed. The equipment must only be put
back into operation when the fault has been repaired.
Every 36 months
Daily
Daily
Daily
Daily
Daily
Daily
Weekly