Equipment Classification - AGFA 5522/100 User Manual

Table of Contents

Advertisement

WS-Manual-001, WS-Manual-T-001, WS-Manual-002, WS-Manual-T-002 | Introduction to DR 400 |

Equipment Classification

Per EN/IEC 60601-1, EN/IEC 60601-2-54, this device is classified as
following:
Table 1: Equipment classification
Class I equipment
Type B equipment
Water ingress
Cleaning
Disinfection
Flammable anesthetics
Operation
Related Links
Cleaning and Disinfecting
Equipment in which protection against electric
shock does not rely on basic insulation only, but in-
cludes a fixed connection to mains power with pro-
tective earth conductor.
A Type B piece of equipment is one that provides a
particular degree of protection against electric
shock particularly regarding allowable leakage cur-
rent and reliability of the protective earth protec-
tion.
IP10
This device does not have protection against in-
gress of water.
See section on cleaning and disinfecting.
See section on cleaning and disinfecting.
This device is not suitable for use in the presence of
a flammable anesthetic mixture with air, or in
presence of a flammable anesthetic mixture with
oxygen or nitrous oxide.
Continuous operation.