62 | WS-Manual-001, WS-Manual-T-001, WS-Manual-002, WS-Manual-T-002 | Remarks for HF-
emission and immunity
from;
to
Table output terminal (x8 24V,
light push button, double expo-
sure protection);
wall stand input terminal
table output terminal (230 V);
wall stand input terminal
table output terminal (AEC);
wall stand input terminal
table output terminal (ground);
wall stand input terminal
Optional
control room (DR Generator
Sync Box 1);
table input terminal (Sync 01)
control room (DR Generator
Sync Box 2);
table input terminal (Sync 02)
control room (DR Generator
Sync Box 1);
wall stand input terminal (Sync
03)
control room (DR Generator
Sync Box 2);
wall stand input terminal (Sync
04)
DX-D Fixed DR Detector or DR
Detector I/O box;
NX workstation
table output terminal Aux.;
control room NX Workstation
table output terminal;
3252C EN 20191206 1610
type;
maximum length
10 x AWG21 (0.5 mm
20 m
3 x AWG18 (1.0 mm
20 m
CAT 5e (SF/UTP);
20 m
2
1 x AWG8 (10 mm
20 m
9 pin sub D (Pin 9 is not
connected);
20 m
9 pin sub D (Pin 9 is not
connected);
20 m
9 pin sub D (Pin 9 is not
connected);
20 m
9 pin sub D (Pin 9 is not
connected);
20 m
CAT 6 SF/UTP;
40 m
Cat 5e;
15 m
01090350F;
remark
mandatory
2
);
mandatory
2
);
shielded
mandatory
);
mandatory
unshielded
unshielded
unshielded
unshielded
shielded (no
connectors
allowed)
shielded
unshielded,
optional