Probe Electronics Wiring; Configuration; Bench Configuration - Genesis ED1 Installation And Operating Manual

Table of Contents

Advertisement

white
red
black
green
63-601 Genesis Models ED1 & ED2 Multiphase Detector

1.4.6 Probe Electronics Wiring

Remove the cover from the probe electronics housing.
Carefully remove the top section of the four-terminal block
within the probe electronics housing.
Carefully feed the 4-position cable through the conduit
entry and connect the four wires in the order of RED,
BLACK, WHITE, and GREEN from left to right as shown:
Refer to wiring diagram within the probe electronics housing.
Insert the terminal block into the socket on the probe elec-
tronics module.
Secure the conduit connection.
Power can now be supplied to the system. Power and
Comm LEDs should be illuminated.
Replace the cover securely onto the probe housing.
1.5

Configuration

Although the Genesis can be delivered pre-configured from
the factory, it can also be easily configured in the shop or at
the installation using the local LCD/Keypad or
PACTw re/DTM. Bench configuration provides a conve-
nient and efficient way to set up the unit before going to
the tank site to complete the installation.
Before starting configuration, collect all operating parame-
ters information. Refer to Section 1.7 for initial
suggestions.
Apply power to the device and follow the step-by-step pro-
cedures below for the menu-driven display. Refer to Section
1.5.2.
Information on configuration using a HART communica-
tor is given in Section 1.6, Configuration Using HART.

1.5.1 Bench Configuration

The Genesis can be easily configured at a test bench by
connecting a 24 VDC (SELV) power supply directly to the
power terminals as shown in the accompanying diagram.
NOTE: When using a HART communicator on Analog Output 1 for
configuration, a minimum 250-ohm line load resistance is
required. Refer to Section 1.6 and your HART communicator
manual for additional information.
NOTE: The detector can be configured without the probe. Disregard
the "Probe Comm Error" diagnostic indicator that will appear.
9