XP Connector in
Main Electronics Base
Align
Push
Connect
6
W RNING! All Genesis probes should be installed such that the
W RNING! Do not disassemble probe when in service and under

1.4.1 Installing the Genesis Main Electronics

Pull
Push
Disconnect

1.4.2 Main Electronics Wiring

Caution: The Genesis operates at voltages of 24 VDC (SELV) (± 10%)
W RNING! Explosion hazard. Do not disconnect equipment unless
maximum level is a minimum of 200 mm (8") below the
flange connection. This may include utilizing a nozzle or
spool piece to raise the probe. Consult factory to ensure
proper installation and operation.
pressure.
The Genesis is offered as standard in a remote configura-
tion, with the base of tank main electronics separated from
the top of tank electronics by up to 30 meters (100 feet).
Using the right angle bracket that is supplied, locate the
main electronics for optimal viewing of the LCD display.
While the housing can be mounted anywhere within the
remote cable distance, this will typically be at the base of
tank. (U-bolts can be utilized with the holes in the right
angle bracket).
1. Ensure the remote cable length ordered is adequate for the
installation.
2. Carefully connect the cable to the main electronics housing
using the "quick-connect" cable as shown at left.
• To connect the cable: Rotate the cable connector to
align the polarity with the housing connector and sim-
ply push on the knurled end of the cable connector. No
tools or tightening required.
• To disconnect the cable: Push on the knurled end of the
cable connector while lightly pulling back on the black
spring loaded ring.
3. Orient the housing for convenient viewing, then secure it to
the bracket using the threaded nut. LCD is rotatable in 90-
degree increments.
with a maximum power draw of 10 Watts.
Wiring connections between the power supply and the
Genesis main electronics should be made using 14–18
AWG shielded twisted pair instrument cable. 4–20 mA
wiring can be accomplished using 18–22 AWG shielded
twisted pair.
power has been switched off or the area is known to be
non-hazardous.
Do not apply power to the system until both the main elec-
tronics and probe electronics wiring is completed.