Diagnostic Indication Table - Genesis ED1 Installation And Operating Manual

Table of Contents

Advertisement

Priority
Indicator Name
Software Error
1
Probe Comm Error
2
Coprocessor Comm
Error
3
TDR Board Error
4
Probe Memory
Error
5
DAC Failure
6
Firmware Conflict
7
Default
Parameters
8
No Top Fiducial
9
No Btm Fiducial
10
Top Safety Zone
Alm
11
Btm Safety Zone
Alm
12
Spare Indicator
13
EoP Above Probe
End
14
EoP Below Probe
End
15
Spare Indicator
16
No Water Signal
17
Initializing
18
Spare Indicator
19
Spare Indicator
20
28
2.3.3 Diagnostic Indicator Table
Below is a listing of the Genesis diagnostic indicators, show-
ing their priority, explanations and recommended remedies.
(Priority 1 is highest priority.)
Default
Explanation
Category
Incorrect instruction sequence or other
Failure
software malfunction.
All AOs
Probe electronics not responding to
Failure
coprocessor.
All AOs
Coprocessor not responding to main
Failure
processor.
All AOs
Probe electronics malfunction
Failure
All AOs
Activated when the memory device in
Failure
probe fails to respond.
All AOs
DAC not responding correctly.
Failure
All AOs
The firmware in the probe electronics is
Failure
not compatible with the firmware in the
All AOs
main electronics.
Some or all parameters reset to default
Failure
values.
All AOs
Top-down waveform missing "fiducial"
Failure
characteristics
All AOs
except Water
Level
Bottom-up waveform missing "fiducial"
FailureWater
characteristics
Level AO
Upper Level signal within Top Safety
FailureUpper
Zone region.
Level AO
Water Level signal within Bottom Safety
FailureWater
Zone region.
Level AO
None
No Effect
End of Probe signal detected closer
Failure
than Probe Length.
All AOs
End of Probe signal detected beyond
Failure
Probe Length
All AOs
None
No Effect
FailureWater
Bottom waveform missing characteris-
tics indicative of water.
Level AO
Function
System initializing, measurements not
yet valid.
Check
none
No Effect
none
No Effect
Remedy (Context Sensitive Help)
Contact Magnetrol Technical Support.
Check wiring / cable connections in
probe electronics compartment.
Contact Magnetrol Technical Support.
Contact Magnetrol Technical Support.
Contact Magnetrol Technical Support.
Contact Magnetrol Technical Support.
Contact Magnetrol Technical Support.
Perform complete Device
Configuration
Contact Magnetrol Technical Support.
Contact Magnetrol Technical Support.
Ensure that liquid cannot reach
Top Blocking Distance.
Ensure that liquid cannot reach
Bottom Blocking Distance.
Check settings:
View Echo Curve.
Check settings:
Probe Length.
View Echo Curve
Standard start-up message. Wait for
up to 10 seconds.
63-601 Genesis Models ED1 & ED2 Multiphase Detector