Initial Configuration/Commissioning - Genesis ED1 Installation And Operating Manual

Table of Contents

Advertisement

20
1.7

Initial Configuration/Commissioning

While the Genesis can be delivered pre-configured for the
specific application, it is relatively simple confirm and/or to
re-configure if necessary.
Refer to Section 1.5 for descriptions of the complete menu
structure and various LCD pushbuttons.
All menus and parameters can be accessed via the LCD user
interface, the DD using a HART configurator, or via
PACTware and the DTM.
The BASIC CONFIG menu is the starting point for the
initial commission process.
Press any key from the HOME Screen
1. Scroll to DEVICE SETUP, then press Enter
2. Scroll to BASIC CONFIG, then press Enter
3. A series of parameters will be displayed:
Level Units
Select the Units of measurement for the level readout:
• Inches
• Feet
• Millimeters
• Centimeters
• Meters
Level Offset
Enter the desired level reading when the liquid is at the end
of the probe. Range is -3 to +6 meters (-10 to +20 feet).
Refer to Section 2.4 for further information. (With default
Level Offset = 0, all measurements are referenced from the
bottom of the probe.)
Upper Dielectric
Enter the Dielectric Constant of the upper process medium.
(Default of 2.1 will suffice for most applications).
Emulsion Top % Water
Enter the desired %Water for the top of the emulsion.
Emulsion Bottom % Water
Enter the desired %Oil for the bottom of the emulsion.
63-601 Genesis Models ED1 & ED2 Multiphase Detector