SECTION 7—FRONT RIGGINGS/FOOTBOARD

ASBA Seat

NOTE: For this procedure, refer to FIGURE 7.7.
1. Remove the two mounting screws, spacers and locknuts that secure the telescoping
front rigging support to the seat frame.
2. Perform one of the following:
• Slide existing telescoping front rigging support to one of three depth positions.
• Remove existing telescoping front rigging.
3. Secure telescoping front rigging at desired depth with existing two mounting screws,
spacers, and locknuts. Securely tighten.
NOTE: The two telescoping front rigging supports can be positioned at different depths depending
on the need of the user.
Spacer
Locknut
FIGURE 7.7 Adjusting/Replacing Telescoping Front Rigging Supports - ASBA Seat

Adjustable ASBA Seat

If the telescoping tubes need to be extended greater than two inches, then the seat
MUST be repositioned rearward to ensure stability - otherwise personal injury and/
or damage to the wheelchair and surrounding property may result.
NOTE: For this procedure, refer to FIGURE 7.8 on page 71.
1. Remove the two cap screws, spacers and threaded blocks securing the telescoping
front tube to the side rail.
2. Perform one of the following:
• Slide existing telescoping front rigging support to one of six depth positions.
• Remove existing telescoping front rigging.
Pronto® M71™
All manuals and user guides at all-guides.com
Mounting
Screws
Spacer
Locknut
WARNING
70
Telescoping Front
Rigging Support
Part No 1118377