2.2 Requirements of Power Supply

UTL8500X series can only be used in following conditions:
Parameters
Frequency
Three-core power cord is provided. Please make sure that the ground wire of
three-phase socket is properly grounded before use.
The instrument comes with 220V/110V power transfer switch. Before connecting to
the power supply, please check and ensure that the switch is in correct position.
250V/0.25A (5x20mm) fuse is selected and installed for the instrument (220V) with a
spare fuse in the fuse case.
In addition, 2 spare fuses of 250V/0.5A are provided for input voltage of AC 110V.
When replacing the fuse, please remove the external power cord first, then open the
fuse slot under the power interface, take out the old fuse and replace it with a new
one, and install the fuse slot back after completion.
Warning: please do not use the damaged power cord to avoid danger. Use the
250V/0.5A fuse when 110V AC power is input.

2.3 Operating Environment

UTL8500X series can only be used in common-temperature and low-condensing zone.
The general environment requirements are listed as follows. During the on-load process,
the speed of the cooling fan will adjust based on the change of the cooling fin's
temperature.
Environment
Operating temperature
Operating humidity
1pc
×
1pc
×
Voltage
Power
AC220V input voltage: 250V/0.25A
Fuse
AC110V input voltage: 250V/0.5A
Users can download the
×
manuals from UNI-T's
official website.
Requirements
AC 220/110 (±10%)V
50/60Hz
50W
Requirements
0°C~40°C
20%~80% (Non-condensing)