and the values of △V and Reg will be detected. During the test, the ON button can be
used to control the load switch.
Operation steps:
1.
In initial interface, select [Mode], [More], [More], and then press [Load Effect] to enter
the setting interface.
2.
Move the cursor to the needed setting position by pulse knob, and press the Enter
button
to change the parameter (the parameter changes from yellow to white)
by keyboard or pulse knob. Press the Enter button
the set value changes from white to yellow.
3.
Other parameters can be modified in a similar way.
4.
Press the ON button, the electronic load starts to load, and the indicator light below
the button lights up. If users need to stop the load, press the ON button again, and the
running indicator light goes out.

4.2.8 OVP Mode

The OVP (overvoltage protection) test function can capture the peak point of the input
voltage and the entire fall process after protection. When the voltage drops to the set
trigger voltage, the load will record the time from the peak point to the trigger point. This
recorded time is the over-voltage protection time of the tested power supply.
Parameter setting in OVP mode:
Trigger voltage
If users want to perform OVP test, they can press More in the menu to select the OVP test
mode, set the voltage of the trigger point, and then press the ON button to start loading.
The overvoltage protection function of the power supply is triggered by increasing the
output voltage of the power supply under test. After the load captures the peak and falling
Figure 4-2-8-1 Interface of OVP Mode
Parameter
Set the voltage of the trigger point
again to confirm. At this time,
Description
37