Environmental Requirements

This instrument is suitable for the following environment:
Indoor use
Pollution degree 2
In operating: altitude lower to 3000 meter;in non-operating: altitude lower to 15000 meter
Operating temperature 0 to +40℃; Storage temperature -20 to﹢70℃ (unless otherwise specified)
In operating, humidity temperature below to +35℃, ≤90% relative humidity;
In non-operating, humidity temperature +35℃ to +40℃, ≤60% relative humidity.
There are ventilation opening on the rear panel and side panel of the instrument. So please keep the air
flowing through the vents of the instrument housing. To prevent excessive dust from blocking the vents,
please clean the instrument housing regularly. The housing is not waterproof, please disconnect the power
supply first and then wipe the housing with a dry cloth or a slightly moistened soft cloth.

Connecting Power Supply

The specification of AC power supply that can input as the following table.
Please use the attached power lead to connect to the power port.
Connecting to service cable
This instrument is a Class I safety product. The supplied power lead has good performance in terms of case
ground. This spectrum analyzer is equipped with a three-prong power cable that meets international safety
standards. It provides good case grounding performance for the specification of your country or region.
Please install AC power cable as follows,
• ensure the power cable is in a good condition;
• leave enough space for connecting the power cord;
• Plug the attached three-prong power cable into a well-grounded power socket.

Electrostatic Protection

Instruments.uni-trend.com
Poor ventilation will cause the device temperature to rise, thus causing damage to this
device. Please keep proper ventilation during use, and regularly check the vents and
fans.
Please take actions to avoid dust or moisture in the air affecting the performance of
this device. Please keep the product surface clean and dry.
The recommended calibration period is one year. Calibration should only be carried out
by qualified personnel.
Voltage Range
100-240VAC (Fluctuations±10%)
100-120VAC (Fluctuations±10%)
Frequency
50/60Hz
400Hz