Step Up - Invacare LiNX REM400 User's Manual Supplement

Control system
Hide thumbs Also See for LiNX REM400:
Table of Contents

Advertisement

External stop switch test
The external stop switch test checks that the external stop
switch is functioning correctly. The test is conducted once
per power cycle when:
the wheelchair is powered up in a latched drive mode
function or
a latched drive mode function is selected following a
non-latched drive mode function.
The external stop switch test is indicated by a screen overlay.
1. Press external stop switch to complete test.
The wheelchair does not drive until the external stop
switch test is completed successfully.
60101834-C

6.6.2 1 Step Up

In this mode, a single drive command (forward or reverse)
causes the wheelchair to accelerate to the maximum drive
speed A of the selected drive screen and then to remain at
that speed for the programmed Latch Drive Timeout period
as long as no further command is given.
Driving
1. Give drive command in desired direction (forward or
reverse).
2. Release drive command.
Wheelchair speed accelerates to maximum drive speed
of the selected drive screen.
Stopping
Use one of the following methods to stop:
Give drive command in opposite direction (a reverse
command when driving forward or a forward command
when driving in reverse)
Press external stop switch
Usage
47

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents