Download Print this page

Kodak HEALTH SCIENCES MINILOADER 2 Plus Low Voltage Stand-Alone Installation Instructions Manual page 27

Hide thumbs Also See for HEALTH SCIENCES MINILOADER 2 Plus Low Voltage Stand-Alone:

Advertisement

INSTALLATION INSTRUCTIONS
15.
Check that the CASSETTE is open the correct distance when the CASSETTE INJECTOR operates. If
necessary adjust PARAMETERS P8 & P9. See the SERVICE MANUAL.
16.
If the MINILOADER is used in a darkroom, the brightness of the DISPLAY can be adjusted by
PARAMETER P10. The brightest DISPLAY is obtained by setting the PARAMETER to 1 and increasing
the number through 2, 3 & 4 progressively dims the DISPLAY. If the DISPLAY is still too bright, fit the RED
FILTER - PART NUMBER 30026275 inside the FRONT PANEL. The brightness of the LIGHT BAR
(TRAFFIC LIGHTS) can be adjusted by R211 on PCB 502. R212 on the same PCB adjusts the BUZZER
volume.
17.
PARAMETERS P11 & P12 control how far the FILM is into the MULTIPLE FILM DETECTOR for
18 x 24 and 24 x 30 respectively when the MULTIPLE FILM DETECTION PAUSE occurs. The leading
edge of the FILM should be 25 - 30 mm into the detector.
18.
Check that the PROCESSOR (if fitted) is brought out of stand-by when a CASSETTE is entered in the
MINILOADER.
19.
Set the date and time on the MICROPROCESSOR (see the THEORY GUIDE). Zero all statistics and
film counters (operator and ESR).
20.
Remove the STEP-BY-STEP SWITCH and the INTERLOCK KEY and re-install all the PANELS.
21.
Load the MINILOADER 2 PLUS with some customer FILM and make some test cycles. Examine the
processed film for artifacts such as scratches, SUCKER marks and light fog. Correct as necessary. If the
MINILOADER is a STAND-ALONE unit, if possible take the film elsewhere to process and do the checks.
22.
Fill in the INSTALLATION REPLY FORM, and note the meter (counter) reading. Make sure the correct
SERVICE CODE is used when feeding back the installation to SCAN.
SERVICE CODE
3419
3420
3421
January 1995
MODEL
STAND-ALONE
M35-M
480RA
Page II-27
II 3420
© Kodak Ltd.

Advertisement

loading