Download Print this page

Kodak HEALTH SCIENCES MINILOADER 2 Plus Low Voltage Stand-Alone Installation Instructions Manual page 18

Hide thumbs Also See for HEALTH SCIENCES MINILOADER 2 Plus Low Voltage Stand-Alone:

Advertisement

II 3420
Change this label to
"CAUTION
RECEIVING MAGAZINE ONLY
CANNOT BE UNLOADED BY A MINILOADER"
when all films are being
collected in single magazine
9.
PHOTOCELL FC7 must be moved so it detects the MAGAZINE FULL PATCH at the HINGE end of
the RECEIVING MAGAZINE. See figure 20.
Figure 20
10.
Unscrew the PHOTOCELL and move it to the opposite end of the mounting bracket. See figure 21.
Remember to adjust the PHOTOCELL to give correct MAGAZINE FULL detection (65 FILMS
maximum) as you complete the installation.
New position of photocell
Figure 21
© Kodak Ltd.
Figure 19
Original position of FC7 for
separate collection of films
INSTALLATION INSTRUCTIONS
24 x 30 label only fixed to magazine when
separate collection method used
New position of FC7 for collection
of all films in one magazine
Page II-18
Reflective patch applied
here only when separate
collection method used.
Original position of FC7
January 1995

Advertisement

loading