INSTALLATION INSTRUCTIONS
II 3420
MODEL STAND-ALONE
Connect the water supply to the WATER INLET. If a
BOTTLE is being used to supply the inlet water, the
RESTRICTOR should be removed. If the water supply
is at mains pressure, the RESTRICTOR MUST be
fitted, otherwise water can splash inside the
MINILOADER when the SOLENOID VALVE opens.
INLET
The drain must be connected. If the water supply is at
mains, the drain should be connected to a fixed customer
DRAIN
drain with a capacity of 1 litre/minute. If the water
supply is from a BOTTLE, a larger size collecting
BOTTLE than the inlet BOTTLE should be used.
IMPORTANT:- If the MINILOADER is in a mobile
unit, or is mounted on a movable STAND, the
customer must be informed that the water must be
turned off and the HUMIDIFIER drained using the
DRAIN VALVE, before the MINILOADER is
moved. If the MINILOADER is not going to be
used for some time, it is advisable to drain the
HUMIDIFIER of water to prevent bacterial growth
in the PAPER ELEMENT.
When the MINILOADER is switched on, the
HUMIDIFIER is also turned on. The DISPLAY will
show the current humidity inside the MINILOADER.
ALTERNATIVE DRAIN POSITIONS
If the humidity inside the MINILOADER is greater
than the preset value, only the DRY FAN will run. If
the humidity is below the preset value, the SOLENOID
VALVE opens, the DRY FAN stops and the WET
FAN starts to increase the humidity.
If the humidity inside the MINILOADER is different
(either higher or lower) from the preset value after 10
minutes, the figures on the DISPLAY will flash. If the
humidity inside is higher than the preset value nothing
can be done, but if the humidity is lower, check the
water supply, FLOAT VALVE and the PAPER
ELEMENT.
RESTRICTOR
For full details of the HUMIDIFIER, see the,
"HUMIDIFIER SERVICE MANUAL" Publication
number XP-3239-2. This Publication includes the
SERVICE DETAILS, THEORY GUIDE, CIRCUIT
DIAGRAMS and PARTS LIST.
Go to the, "INSTALLATION CHECK LIST" on
page II-24 of this PUBLICATION.
January 1995
Page II-23
© Kodak Ltd.