II 3420
Figure 13.
New clear panel
(30026252)
Countersunk
screw 6
22.
Position the MINILOADER and STAND in front of the M35-M. Pull the Miniloader over the
PROCESSOR (lifting it slightly to clear the STOPS). Fasten the CABLE RETAINING BLOCK to the
PROCESSOR REAR PANEL using the SCREWS provided. See FIGURE 14. Check that the CABLE
is not strained when the MINILOADER is pulled forward to its STOPS. Lay the LIGHT BAFFLE in
position on the top of the PROCESSOR. Plug in the two FLYING LEADS for PHOTOCELL FC9 and
STAND-BY OVER-RIDE. Pull the MINILOADER into its fully docked position and ensure the relative
heights of the MINILOADER and PROCESSOR are correct. Adjust if necessary.
Figure 14
23.
Complete the installation of the M35-M PROCESSOR, using the M35-M INSTALLATION
MANUAL. IMPORTANT, THE M35-M MUST BE INSTALLED IN ACCORDANCE WITH
THE M35-M INSTALLATION MANUAL AND ADDITIONALLY IN ACCORDANCE WITH
LOCAL PRACTICES. Then go to, "INSTALLING THE HUMIDIFIER" on page II-22 of this
PUBLICATION.
© Kodak Ltd.
Countersunk
screws 4 & 5
Countersunk
screw 7
Page II-12
INSTALLATION INSTRUCTIONS
New metal plate
(30015562)
Countersunk screw 8
(no cup washer here)
Cable retaining
block
Light baffle
Flying
leads
January 1995