3.
Unsolder the pins of the image sensor to detach the PA-418 board (channel G).
PA-418 board (channel B)
Note
• Connect the flexible flat cable to the connector on the PA-418 board, before soldering the PA-418 board.
• When installing the OHB bracket, tighten screws (a) to (d) in alphabetical order.
4.
Install the removed parts by reversing the steps of removal.
5.
Perform "Actions to Be Taken during Board Replacement and after Board Replacement/Repair". (Refer to
Notes on Replacement of Parts and Circuit
HSC300RF/HSC100RF/HSC300R/HSC100R
Image sensor
PA-418 board (channel R)
Board")
OHB bracket
(a)
PSW2 x 5
(b)
PSW2 x 5
CN1
PA-418 board (channel G)
(d)
PSW2 x 5
(c)
PSW2 x 5
Flexible flat cable
"1-4.
2-17