Maintenance
Simple Repairs

Reinstalling the Pump Head Assembly

When
When reassembling the pump
Tools required
• 4-mm hexagonal key
Parts required
#
79846-65501
1 Apply a small amount of grease on the back of the pistons.
2 Slide the Pumphead Assembly onto the Pumpdrive.
3 Using a 4 mm hexagonal key tighten the Pumphead screws stepwise with
4 Reconnect the capillaries, tubing and the Active Inlet Valve cable to the
5 Reinstall the front cover.
100
Description
Pump head grease
increasing torque.
connector.
1200 Series Quarternary Pump User Manual