Advertisement

7
Maintenance
Simple Repairs
5 Using a pair of tweezers or a toothpick remove the frit.
6 Place a new frit into the purge valve with the orientation of the frit as
7 Reinstall the cap with the gold seal.
Before reinstallation always check the gold seal. A deformed seal should be exchanged.
N O T E
8 Insert the purge valve into the pump head and locate the pump outlet
86
Figure 17
Purge Valve Parts
shown above.
capillary and the waste tube as shown in
Figure 18
on page 87.
1200 Series Quarternary Pump User Manual