Optimizing Performance

Optimize the Compressibility Compensation Setting

Optimize the Compressibility Compensation Setting

The compressibility compensation default setting is 100 × 10
quaternary pump. This setting represents an average value. Under normal
conditions the default setting reduces the pressure pulsation to values (below
1% of system pressure) that will be sufficient for most applications and for all
gradient analyses. For applications using sensitive detectors, the
compressibility settings can be optimized by using the values for the various
solvents described in
the compressibility tables, when using isocratic mixtures of solvents and if the
default settings are not sufficient for your application the following procedure
can be used to optimize the compressibility settings.
When using mixtures of solvents it is not possible to calculate the compressibility of the
N O T E
mixture by interpolating the compressibility values of the pure solvents used in that mixture
or by applying any other calculation. In these cases the following empirical procedure has
to be applied to optimize your compressibility setting.
1 Start the pump with the required flow rate.
2 Before starting the optimization procedure, the flow must be stable. Use
3 Your pump must be connected to a control software (e.g. ChemStation,
4 Start the recording device with the plot mode.
5 Starting with a compressibility setting of 10 × 10
60
Table 9
degassed solvent only.
EZChrom, OL, etc.) or handheld controller with which the pressure and
%-ripple can be monitored, otherwhise connect a signal cable between the
pressure output of the pump and a recording device (for example, 339X
integrator) and set parameters.
Zero 50%
Att 2^3 Chart
Speed 10 cm/min
in steps of 10. Re-zero the integrator as required. The compressibility
compensation setting that generates the smallest pressure ripple is the
optimum value for your solvent composition.