Performance Specifications

Table 4
Type
Hydraulic system
Setable flow range
Flow range
Flow precision
Flow accuracy
Pressure
Pressure pulsation
Compressibility
compensation
Recommended pH
range
Gradient formation
Composition Range
Composition Precision
Control and data
evaluation
Communications
26
Performance Specification Agilent 1200 Series Quaternary Pump
Specification
Dual plunger in series pump with proprietary servo-controlled variable
stroke drive, floating plungers and active inlet valve
0.001 – 10 ml/min, in 0.001 ml/min increments
0.2 – 10.0 ml/min
£ 0.07% RSD, or £ 0.02 min SD whatever is greater, based on retention
time at constant room temperature
± 1% or 10 µl/min whatever is greater
Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) up to 5 ml/min
Operating range 0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) up to 10 ml/min
< 2 % amplitude (typically < 1 %), at 1 ml/min isopropanol, at all
pressures > 1 MPa (10 bar)
User-selectable, based on mobile phase compressibility
1.0 – 12.5, solvents with pH < 2.3 should not contain acids which attack
stainless steel
Low pressure quaternary mixing/gradient capability using proprietary
high-speed proportioning valve Delay volume 800 – 1100 µl, dependent
on back pressure
0 – 95 % or 5 – 100 %, user selectable
< 0.2 % RSD, at 0.2 and 1 ml/min
Agilent control software (e.g. ChemStation, EZChrom, OL, etc.)
Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready,
start, stop and shut-down signals, LAN optional